Macro-channel stacked array laser DSHDV-808ag-800W-D16 DSHDV-755&808&1064ag-800W-D16

. main features:

Macro channel has low water quality requirements, gold and tin packaging, high reliability, and peak value power

. applied range:

medical treatment

. Detailed parameters:

 

 

Parameter (25℃)

 

unit

 

HD-8280-D16/G16

 

HD-M280-D16/G16

Optical parameters

 

 

 

The central wavelength, λ c

nm

808

755/808/1064

The central wavelength deviation

nm

±10

±10

output power

W

800

800

Barbar number

twig

16

16

Bar spacing

mm

2

2

Fast-axis divergence angle (95%)

degree

<50

<50

Slow-axis divergence Angle (95%)

degree

<10

<10

spot size

mm

10x32.5

10x32.5

Electrical parameters

 

 

 

Operating current Iop

A

55

55

Threshold current I th

A

<13

<13

Working voltage, Vop

V

≤32

≤32

service frequency

Hz

110

110

pulse width

ms

≤200

≤200

duty cycle

-

≤20%

≤20%

enviromental parameter

 

 

 

Work water temperature

20-30

20-30

storage temperature

0-55

0-55

water gage

MPa

≤0.6

≤0.6

Water flow

L/min

2.5-4

2.5-4

 

. boundary condition:

 

HD-800W  Energy Table

Single  Energy(J)

Frequency (HZ )

Iop

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

Pulse

Width (m

s)

10

8.8

8.8

8.8

8.8

8.8

8.8

8.8

8.8

8.8

8.8

 

55A

20

17.6

17.6

17.6

17.6

17.6

17.6

17.6

17.6

17.6

17.6

 

50A

30

26.4

26.4

26.4

26.4

26.4

26.4

26.4

26.4

26.4

26.4

 

45A

40

35.2

35.2

35.2

35.2

35.2

35.2

35.2

35.2

35.2

35.2

 

 

50

40

30

30

30

30

30

30

30

 

 

 

 

Cooling  water T  =25±3℃;

Cooling  water flow  rate:

2-4L/min;

60

48

36

36

36

36

36

 

 

 

 

70

56

42

42

42

42

 

 

 

 

 

80

64

48

48

48

48

 

 

 

 

 

90

72

54

54

54

 

 

 

 

 

 

100

80

60

60

60

 

 

 

 

 

 

200

144

108

 

 

 

 

 

 

 

 

300

216

 

 

 

 

 

 

 

 

 

400

288

 

 

 

 

 

 

 

 

 

. Notes for use:

  1. Make sure that the circulating water is connected before the laser is energized.
  2. Please avoid using and storing them in the condensation environment.
  3. Over the rated working current will shorten the service life of the device.
  4. Working at higher temperature will reduce the power and shorten the service life of the device.
  5. When the laser is working, avoid laser irradiation on the eyes.